Standard Testing Capability

 • Routine Spirometry (SVC, ERV, IC etc.)
• Flow Volume Loops (FVC, FEV1, PEFR, FEF25-75 etc.)
• Maximum Voluntary Ventilation (MVV, FEV1 x 35 etc.)
• Volume of Thoracic Gas (VTG, FRC, IC, TLC, RV, RV/TLC etc.)
• Airways Resistance (Raw, sGaw, sRaw, Gaw, VTG, FRC, TLC etc.)
• Bronchial Provocation (PD20, PC20 – any and all parameters)
• Manual Entry (ABG’s, Oximetry and Exercise)

Optional Testing Capability

• Single Breath Diffusion (DLCO, VA, DL/VA etc.)
• Maximum Inspired and Expired Pressures (MIP/MEP)

View Brochure

The Experience

 Working with the Body Box is the ultimate pulmonary function testing experience. The instrument cabins are superbly crafted using the finest aluminum extrusion techniques and precision engineering of the glass doors, seals, and patient valving. The products are hand crafted with the pride and skill that is the heritage of Europe’s finest engineering workshops. The all-round glass panel enclosure provides space, comfort, and reassurance to subjects normally wary of being in a closed space.

With our heritage of working with research centers and teaching hospitals, the Body Box offers a wide range of testing capability. The design provides flexibility in preferences of testing protocols. For example, the single breath diffusion option allows for testing with an inspiratory demand valve, or with an inspiratory bag reservoir. It can offer fast response infrared gas analysis for rapid DLCO measurements or traditional techniques using fuel cell CO technology.

The genius and craft of the valving design makes such versatility possible. The plethysmograph cabin, electronics, and patient valve assembly have all been carefully designed to suit tests on subjects ranging from very small children to large adults. The patient valve can be simply adjusted to ensure comfort and proper testing posture. The shutter valve is quiet and efficient. It utilizes compressed gas to gain the efficiency and reliability of pneumatic sealing over thousands of operations.

View Brochure

Comfort

 Patients can easily exit the cabin should they feel uncomfortable, and there is two-way communication for reassurance and testing instruction. The ComPAS software design focuses technician attention on the patient. The tests are carried out quickly with the least inconvenience or patient concern. The cabin interior is a bright, comfortable, and reassuring environment!

View Brochure

ComPas & Body Box

 Too many pulmonary function instruments in today’s competitive market are designed with obsolescence in mind. Not so with any of the Morgan Scientific or MediSoft products! It is not unusual for our systems to operate for 15 years or more. The cost of maintaining our magnificently engineered Body Box is far less than other manufacturers.

Not only is the cost of service negligible, but the costs of testing supplies are also far less than alternative systems.

For DLCO studies, only a single cylinder of diffusion gas is required. A typical ‘H’ cylinder will allow approximately 300 complete PFT studies to be completed.

The ComPAS software matches the excellence of the Body Box hardware. For many customers, it is the extension of our program first introduced in 1980. Throughout the evolution of Morgan Scientific software, each new edition has provided full conversion of patient data.

ComPAS guides each test maneuver with clear, helpful screens. From the morning’s calibration to complete testing, the software guides, instructs, and assists in assurance of high quality testing. Using ComPAS gives the operator a wise, helpful colleague sitting by their side!

View Brochure

 

TECHINCAL SPECIFICATIONS
General Specifications
Patient Manifold
Body Box Dimensions Pneumotachograph Lilly Style
(H x W x D) Inches: 69.3 x 34.25 x 27.95 Range: 0.001 to 15 L/sec
cm: 176 x 87 x 71 Resistance: 0.4 cmH2O/L/sec
Body Box Weight 286 lbs (130 kgs) Accuracy: < 3%
Internal Volume 960 Liters Linearity: Look-up table
Patient Chair Pneumatic adjustable Calibration: Using a 3L syringe
Body Box Door Closing: Simple external & internal lever mechanism Pneumotach Cleaning: Easily disassembles for cold sterilization.
Power Requirements 100 to 250 VAC 60 Hz Patient Valve Pneumatic operation at 50psi
Power Consumption Approx. 100 VA Deadspace: < 100ml
Warm-up Time Approx. 15 min Valve Support Arm Fully adjustable
Pressure Transducers
PC Interface
Box Pressure Mouth Pressure Flow Semiconductors protected from overload Serial Interface 12 & 14 bit digitization.
Sensitivity: Box P Mouth P Mouth P (MIP/MEP) Flow Sampling Frequency: 100 and 500 Hz /channel
Resolution: Box P Mouth P and Flow   0.0001 cmH2O0.001 cmH2O Serial Transmission: 115,200 baud
Linearity: < 0.1% Isolation circuit: System fully isolated by optical infrared
Accuracy: < 0.01% Computer PC Compatible
Calibration Calibration standards are supplied Operating System: Windows® 98.ME, XP,NT, or 2000
Box P: 30ml integrated pump Recommended RAM: 512 MB
Mouth P: Pressure manometer Recommended Speed: 1 GHz or greater
Flow: 3 Liter Cal Syringe
Standard Gas Analyzers
Optional Gas Analyzers
Helium Analyzer Thermal conductivity Combination Carbon Monoxide & Methane Analyzer Rotating filter high-speed infrared spectrometer
Range: Accuracy: Response Time: 0 to 15% He +/- 1% (2 ­ 14% He) Fixed to 30 seconds Range: Accuracy: Response Time: 0 to 0.350% 50ppm < 0.1 secs (10 – 90% FS)
Oxygen Analyzer Galvanic Cell Oxygen Analyzer Paramagnetic
Range: Accuracy: Response Time:   0 to 100% O2 +/- 1% (0 ­ 100% O2) Fixed to 30 seconds Range: Accuracy: Response Time: 0 to 100% O2 +/- 1% (0 ­ 100% O2) 2 secs (10 – 90% O2)
Carbon Monoxide Chemical Fuel Cell
Range: Accuracy: Response Time: 0 to 0.500% CO +/- 2% (0 ­ 0.3% CO) Fixed 30 seconds
ComPas

Learn MoreView Brochure
TransAir
Learn MoreView Brochure

 